Specifications

Available in a number of models, the Ninjo™ system comprises of a series of high pressure water cycles delivered through strategically located fixed and rotating nozzles, followed by a steam phase. 

Key features

  • Easy to use - simply load, close and activate programme. 
  • 9+2 - system includes nine fixed and two rotating nozzles to meet the highest cleaning requirements.
  • Pipe System Disinfection (PSD) provides rapid disinfection through all fixed and rotating nozzles, disinfecting the entire pipe and nozzle system with every cycle.
  • Smooth, hygienic chamber means no folds or pockets.
  • Ao value display - disinfection levels displayed, confirming that the appropriate level of disinfection has been reached. 
  • Remaining time displayed in an easy-to-follow cycle sequence.
  • Two programmes available: economy and normal, with an additional three available upon request.
  • Locking system during cycle, securing staff safety and ensuring a complete cleaning and disinfection process.
  • Optional drying/cooling fan for rapid and hygienic dry and cooling of recepticles with HEPA filtered (H13) air.
  • Meets and exceeds the stringent cleaning and disinfection requirements of EN ISO15883 (CAN/CSA Z15883).
  • Choice of polymeric material or stainless steel for chamber or outer casing.
  • Environmentally friendly – low consumption of water and energy, recyclable material.
  • Wide range of holders for different receptacles.
  • Lockable storage for detergent containers.
  • Easy maintenance and service.
NINJO - TECHNICAL DATA
UTILITY REQUIREMENTS
Utility  Unit connection Supply Pressure requirments Flow
requirement
Comments
Cold water (CW) 15 mm (G½ “) 100-800 kPa (1-8 bar, 14-116 PSi) 20 L/min See Notes 1 and 3
Hot water (HW)  15 mm (G½ “) 100-800 kPa (1-8 bar, 14-116 PSi) 20 L/min See Notes 2 and 3
Drain (D)  Ø 90 mm alt.
Ø 110 mm
N/A Peak 1 L/sec See Notes 3 and 4
Electrical
Voltage Recommended service disconnect Full load, Amps See notes 5 and 6
L1, 230 VAC, 1N+PE, 50 Hz  Max. 3,0 kW, 1 x Fuse C 16 A 13 A  
L5, 240 VAC, 2+PE, 60 Hz Max .3,3 kW, 2 x Fuse C 15 A 13,6 A  
L9, 208 VAC, 2+PE, 60 Hz Max. 3,0 kW, 2x Fuse C 15 A 14,4 A  
L14, 200 VAC, 2+PE, 50 Hz  Max .3,0 kW, 2x Fuse C 16 A 15 A  
L18, 200 VAC, 2+PE, 60 Hz Max. 3,0 kW, 2x Fuse C 16 A 15 A  
L22, 230 VAC, 1N+PE, 60 Hz  Max. 3,0 kW, 1x Fuse C 16 A 13 A  
L28, 240 VAC, 1N+PE, 50 Hz  Max. 3,3 kW, 1x Fuse C 16 A 13,6 A  
OPERATING CONDITIONS          
Room temperature  5-40°C (41-104F)         
Air humidity  < 80% at 31°C (87° F)         
Max surface temperature  40° C (104°F)         
Noise level          
Octave band (Hz)       125  250  500 1000   2000 4000  8000    LwA  LwAFmax The sound power level as linear
octave band levels are presented
as A-leveled sound power level,
both LwA and LwAFmax. 
Correction (Kok)       57  61  58  55  53  48  47   61  75
Estimated LpA levels for Ninjo located in a 70 m³ room, db ref 20 μPa.          
Type of room     Description     Operator location *    3 m from washer   The calculated sound power levels
are different LpA levels in different
types of rooms. With bigger room
volume the noise level is reduced
and vice versa. Here are three
examples of expected LpA levels.
* See Note 7
   
Hard    All walls from concrete etc no absorbing elements.
   
 60   60 
Normal    Some sound absorbing elements like furniture and textiles.
   
 56   55 
Subdeued  Some absorbing elements in ceiling and textiles.
   
 53     51 
Water consumption          
Economy program    11 liter ± 10 %*   

 * if option cooling/drying fan is selected for the cooling phase. If internal cooling is

selected, additionally 3 liters
± 1 liter will be added to the stated consumption.  

Normal program    18 liter ± 10 %*    
Chamber size and volume
Stainless steel 565x410x290, 67 liter Polymeric 535x400x300, 64 liter
Outer dimensions
  Freestanding Under counter  
Width 450 mm (17” 11/16) 450 mm (17” 11/16) See installation drawing in subsection installation.
Depth  580 mm (22” 3/4) 580 mm (22” 3/4)
Height  1320 mm (51” 7/8) 870 mm (34” 3/16)
900 mm (35” 7/16) with lid
Weight 90 kg/200 lb 70 kg/155 lb

Notes
1) Cold water recommended quality = potable water accordingly to WHO guidelines, temp. <20°C (<68°F).
2) Hot water recommended quality = potable water accordingly to WHO guidelines, temp. 45-60°C (113-140°F).
3) It shall be the Customers responsibility to eliminate water hammer conditions if they occur in the service piping.
4) It shall be the Customers responsibility to provide a properly sixed and located drainage system in accordance with the national
plumbing code. Consider that other equipment is tied to drain to eliminate slow drainage or backup.
5) It shall be the Customers responsibility to complete all electrical connections using properly sized wiring in accordance with the
national electrical code.
6) It shall be the Customers responsibility to install an approved and fused disconnect switch accordingly to national code in all
electrical supply lines at the washer disinfector location.
7) LpA levels are measured to EN ISO 3747:2000, 1 meter in front of the washer and 1,5 meter above the floor level.

Systems available

  • Freestanding  
  • Under-counter

Accessories

  • ArjoHuntleigh bedpan with lid
  • Drainage floor Ø 90 mm and 110 mm (3 1/2" and Ø 4 1/4”) 
  • Drainage wall Ø 90 mm and 110 mm (3 1/2" and 4 1/4”)
  • Valves and filter
  • Getinge Clean Flusher™ Detergent, 5L canister