Specifications

A top loading flusher disinfector, the Amigo™ is exceptionally easy-to-use and has an integrated slop basin.  Designed to improve sluice room efficiency and the working environment, the system saves time, space and money.

AMIGO - TECHNICAL DATA
UTILITY REQUIREMENTS
Utility   Unit connection Supply Pressure requirments Flow requirement Comments
Cold water (CW) 15 (½ “) mm  100 - 800 kPa 20 L/min See Notes 1 and 4
Hot water (HW) 15 (½ “) mm   100 - 800 kPa 20 L/min See Notes 2 and 4
Steam 15 (½ “) mm  30-300 kPa 0,3 kg/process See Note 4
Drain (D) Ø 110/90mm  N/A N/A See Notes 3 and 5
Electrical (E) Recommended service disconnect Full load amps  
Electrical Heating:
240 V, 3+PE, 60 Hz......................................D15.......................................... 16A
240 V, 2+PE, 60 Hz......................................D20.......................................... 18A
240 V, 1N+PE, 50 Hz...................................D20........................................... 18A
230/400 V, 3N+PE, 60 Hz.............................C10........................................... 10A
230/400 V, 3N+PE, 50 Hz.............................C10........................................... 10A
230 V, 3+PE, 60 Hz.....................................C16A......................................... 16A
230 V, 1N+PE, 60 Hz...................................C20........................................... 19A
230 V, 3+PE, 50 Hz.....................................C16........................................... 15A
230 V, 1N+PE, 50 Hz...................................C20........................................... 19A
208 V, 3+PE, 60 Hz.....................................D22........................................... 18A
208 V, 2+PE, 60 Hz.....................................D22........................................... 21A
200 V, 3+PE, 60 Hz.....................................D20........................................... 18A
200 V, 2+PE, 60 Hz.....................................D25........................................... 22A
200 V, 3+PE, 50 Hz.....................................D20........................................... 18A
200 V, 2+PE, 50 Hz.....................................D25........................................... 22A
See Notes 6 and 7
Steam Heating:
240 V, 2+PE, 60 Hz.....................................D10............................................ 6A
230 V, 1N+PE, 60 Hz...................................C10............................................ 6A
230 V, 1N+PE, 50Hz....................................C10............................................ 5A
208 V, 2+PE, 60 Hz.....................................D10............................................ 6A
200 V, 2+PE, 50 Hz.....................................D10............................................ 5A
200 V, 2+PE, 60 Hz.....................................D10............................................ 6A
See Notes 6 and 7
OPERATING CONDITIONS
Room temperature 5-40°C (41-104F)  
Air humidity < 80% at 31°C (87F)  
Max surface temperature 40° C  
Noise level
Calculated sound power level Lw for washer, dB ref 1 pW. Disinfection is excluded in the measurement
Octave band (Hz)       125 250 500 1000 2000 4000 8000  LwA LwAFmax The sound power level as
linear octave band levels are
presented as A-leveled sound
power level, both LwA and
LwAFmax.
Correction (Kok)     60 56  56  58 58 56 57 64 80
Noise level
Estimated LpA levels for Amigo located in a 70 m³ room, db ref 20 μPa.
Type of room 
  
Description Operator location * 3 m from washer The calculated sound power levels are different LpA levels in different types of rooms. With bigger room volume the noise level is reduced and vice versa. Here are three examples of expected LpA levels.
* See Note 8
Hard All walls from concrete etc no absorbing elements 63 63
Normal Some sound absorbing elements like furniture and textiles. 59 58 The calculated sound power levels are different LpA levels in different types of rooms. With bigger room volume the noise level is reduced and vice versa. Here are three examples of expected LpA levels.
* See Note 8
Subdeued Some absorbing elements in celing and textiles 56  54
Water consumption (without cooling)
Economy 18 liter ± 10 % Cooling approx. 3 liter
Normal 29 liter ± 10 %
Intensive 36 liter ± 10 %
WEIGHTS AND DIMENSIONS
Outer Dimensions
Width 590 mm
Depth 660 mm
Height 1000 -1060 mm
Weight  85 kg
Effective Chamber Dimensions 
Width 465 mm
Depth 400 mm
Height 400 mm
Volume 70 L
Material of construction Stainless steel EN 1,4404, ASTM 316 L

Notes:

1) CW recommend quality = drinking water (see Installation Manual) and temperature <20°C (<68F).

2) HW recommend quality = drinking water (see Installation Manual) and temperature 45-60°C (113-140F).

3) Without drain cooling is the temperature up to 90°C (194F). With drain cooling 60°C (164F).

4) It shall be the Users responsibility to eliminate water or steam hammer conditions should they occur in the

service piping.

5) It shall be the customers responsibility to provide a properly sized and located drainage system in accordance with the

national plumbing code. Consider that other equipment is tied to train to eliminate slow drainage or backup.

6) It shall be the User responsibility to complete all electrical connections using properly sized wiring in

accordance with the national electrical code.

7) It shall be the customers responsibility to proved a fused disconnect switch in all electrical supply lines at the washer

location.

8) LpA levels are valid when washer is built in. Measured 1 meter in front of the washer and 1.5 meter above the

floor level.

 

Top loading, slanted flush chamber 

  • Top loading design supports ergonomic working methods.
  • Slanted front for easy loading and emptying.
  • Deep-drawn chamber design minimises risk of bacterial residues.  

Integrated slop basin 

  • Button press-activated rim flush.
  • Efficient emptying of receptacles.
  • Effective post-process chamber disinfection.  

User-friendly control system 

  • LCD display and control panel presenting clear text and process indicators.  

Easy programme selection 

  • Memory available for up to 25 programmes.
  • Five programmable buttons activate the most common programmes.
  • The most frequently used programme can be pedal activated.  

Wall-mounted or freestanding option

  • Easy under-machine access for floor cleaning.
  • Depressing the pedal opens and closes the lid, preventing contamination.
  • Pedal action can be used to start the most frequently used programme, making it efficient and hygienic.

Systems available

  • Wall-mounted
  • Freestanding